A preview of the papers appearing in the June 2007 issue of the Journal of the SID. To obtain access to these articles on-line, please go to www.sid.org

Edited by Aris Silzars


Medical displays in the healthcare system

Adi Abileah 
Planar Systems

Abstract — As the healthcare system changes and progresses, the need for different types of high-performing displays is also evolving. There are three categories of displays: (a) embedded (as part of life-saving devices), (b) informative (for patient data and history, and managing workflow), and (c) imaging (high performing for diagnosis). The challenges of AMLCDs, which are the display of choice at the moment mainly in digital imaging, will be discussed. These challenges include very high resolution, high brightness, and wide viewing angle. The current performance of AMLCDs and the areas which they need to improve will be reviewed. A brief summary of the standards used to specify medical (diagnostic) displays will follow. A look into the future will predict the role of displays in hospitals.

Digital imaging (DI) is a more-demanding area of the healthcare system. Based on the imaging device and viewing the results on the display, a critical diagnosis is made. In many instances it can be a question of life or death. Therefore, the responsibility of the display specifications and performance are very important. In the case of X-ray imaging, the quality of the X-ray sensors, electronics, and the display, all as a system, has to match or exceed the performance of X-ray films. This is not trivial. Therefore, very-high-resolution displays need to be used.

p34_tif

FIGURE 8 — A 4-Mpixel display showing typical information.

Image display in endoscopic surgery

Christopher S. Muratore
Beth A. Ryder
Francois I. Luks

Brown Medical School

Abstract — Advances in the technology of optical displays have changed the way surgeons are able to manage different illnesses. Minimally invasive surgery encompasses a wide range of endoscopic procedures, whereby the body cavity (abdomen, thorax, gastrointestinal tract, and joint spaces) is accessed through small incisions and the use of telescopes and fine, long instruments. These techniques have rapidly gained in popularity during the last decades, as patients are experiencing less discomfort after surgery. Visualization of the operative field requires optimal image capture, processing, and display. The introduction of charge-coupled devices has enabled surgeons to view the operative field on a video monitor, allowing ever-more-complex operations to be performed endoscopically. However, limitations include loss of 3-D perception and tactile sense, poor ergonomics, often suboptimal positioning of the image display, and image quality that is too dependent on outside influences. These limitations, and possible solutions, are addressed as is the "ideal" display system for endoscopic surgery.

Recently, flat-panel liquid-crystal-display (LCD) monitors have provided an improved image quality at a reduced price, and many operating rooms are now replacing CRT monitors with flat screens. Their light weight, versatility, and small footprint allow the use of multiple screens placed near the operating table. While LCD monitors produce the best image and have a greater portability than CRT monitors, the positioning of the screen is still limited by a number of factors, including the need to keep the immediate surroundings of the operative field sterile and the presence of a large number of instruments, instrument tables, cables, and other equipment that add to the clutter of the endoscopic operating room (Fig. 4).

p35a_tif

FIGURE 4 — Typical laparoscopic operating room set-up. Note the multitude of image displays: Two flat-screen laparoscopic monitors (A and D, off the laparoscopic "tower"), vital signs and anesthesia parameters display (B), patient data entry and display system (C), and drug-information display (E).

 


Improving information quality of MR brain images by fully automatic and robust image analysis methods

Ahmet Ekin
Radu Jasinschi
Jeroen van der Grond
Mark van Buchem

Philips Research

Abstract — The medical experts are expected to make diagnosis on an ever-increasing amount of data. Although the intention with the technological advances in imaging devices, such as MRI (magnetic resonance imaging), is to improve the medical practice, the increased amount of data may make the medical expert's task even harder and could arguably be counter-productive. A more desirable situation for the medical expert is the improvement in the clinical relevance of the image content and a faster, preferably automated, access to the most relevant part of the existing large amount of data. To this effect, the concept of information quality is introduced, and fully automatic, fast, and robust image analysis methods that improve it are proposed. As a guideline for a general framework, these methods are classified by the type of information they can improve: geometric, structural (sub-organ), and organizational. The algorithmic contributions are in the domain of MR brain image analysis. These methods improve the information quality not only for the visualization and the diagnosis by the expert, but also for the quantification by other automated methods. The robustness of these algorithms has been extensively tested at the Leiden University Medical Center over a large database of 550 patients.

The geometry of the brain is utilized in various ways to make diagnoses. For example, a healthy brain exhibits a rough bilateral symmetry with respect to the interhemispheric fissure that bisects the brain. This fissure is commonly referred to as the anatomical mid-sagittal plane (MSP) and is shown in Fig. 1. The radiologists use the expected symmetry about the MSP to detect abnormalities. Because the head of the patient may be in an arbitrary position during scanning, the evaluation of hemispheric symmetry from the raw data is difficult.

p35b_tif

FIGURE 1 — The mid-sagittal plane location for the brain MR slice on the left is shown as a white line on the right.

Evaluation of high-resolution and mobile display systems for digital radiology in dark and bright environments using human and computational observers

Rebecca Vogel
Anindita Saha
Kish Chakrabarti
Aldo Badano

Food and Drug Administration

Abstract — As digital display systems replace film traditionally used for reading radiographic images, resource-intensive acceptance testing must be performed to ensure that quality meets and maintains desired specifications. If machine observers can replace human readers, whose performances are highly variable, the results will be more consistent and less costly. To be effective, however, the automated observers must track human performance. An approach for a model observer, validated with human readers, for the evaluation of the visibility of low-contrast small targets in high-resolution and mobile displays under different ambient illumination, will be described. The displays were tested using CDMAM-like digital phantoms containing disks of varying diameters and contrasts on a flat background. For this task, we find the best indicator of display performance to be the display's ability to represent small luminance contrast, not resolution or pixel size. The results confirm that high-resolution systems perform better under low illumination while illuminance has a minor impact on the mobile-display performance. Finally, the results show that the machine observer tracks the performance of human readers. Machine observers with proper validation can replace humans in the acceptance testing procedures, saving the testers both time and money.

Seven display systems were used in this study: four high-resolution monitors and three mobile displays. Of the four high-resolution displays, two were 5-Mpixel gray-scale monitors from different manufacturers, one was a 3-Mpixel color monitor, and one was a 1-Mpixel color monitor commercialized as medical-display monitors for use in radiology and other imaging fields such as reconnaissance, advanced graphics, and geospatial and military applications.

p36a_tif

FIGURE 5 — The set-up of the acquisition of images for the machine observer.

Clinical applications for stereoscopic 3-D displays

David J. Getty
Patrick J. Green

BBN Technologies

Abstract — Stereoscopic 3-D digital imaging holds the promise of improving the detection, diagnosis, and treatment of disease as well as enhancing the training and preparation of medical professionals through use of stereoscopic 3-D displays in concert with the many volumetric visualization techniques/modalities developed in recent years. While so-called 3-D graphics have improved the state of computer visualization in general, 3-D displays make full use of the human-visual perception, and thus can provide critical insight in complex computer-generated and video 3-D data. The stereo 3-D applications reviewed in this paper include screening of breast cancer and diabetic retinopathy, visualization for minimally invasive surgery, and the teaching of anatomy. Also included is a discussion of groundbreaking results from a stereo digital mammography clinical trial under way at Emory University.

A clinical trial of stereoscopic digital mammography versus standard digital mammography in a screening setting is currently under way. The interim results from the trial are striking. In the current case sample, stereo mammography has reduced false-negative readings by 44% (27 false-negative readings by standard mammography, compared to only 15 by stereo mammography). Equally impressive, stereo mammography has reduced false-positive lesion detections in the current sample by 37% (68 false positive detections by standard mammography compared to only 43 for stereo mammography).

 

p36b_tif

FIGURE 8 — Use of a 5-Mpixel Planar StereoMirror™ monitor in the Stereo Digital Mammography clinical trial.

 


Influence of 8-bit vs. 11-bit digital displays on observer performance and visual search: A multi-center evaluation

Elizabeth A. Krupinski
Khan Siddiqui
Eliot Siegel
Rasu Shrestha
Edward Grant
Hans Roehrig
Jiahua Fan

University of Arizona

Abstract — Medical-grade monochrome monitors typically display 8 bits of data. This study determined if 11-bit displays could improve observer performance and decrease use of window/level. 8- and 11-bit displays from three manufacturers were used at three sites. Six radiologists at each site viewed 100 DR chest images (half with a pulmonary nodule) on both displays. Decisions, confidence, nodule location, viewing time, and window/level use were recorded. There was no significant difference in ROC Az as a function of bit depth. The average Az with 8 bits was 0.8284 and with 11 bits was 0.8253. There was a significant difference in viewing time favoring the 11-bit displays. Window/level use did not differ. Eye position was recorded on a subset of images at one site. Cumulative dwell times for each decision category were lower with the 11-bit than with the 8-bit display. When tested with t-tests for paired observations, the TP (t =1.452, p = 0.1507), FN (t = 0.050, p = 0.9609), and FP (t = 0.042, p = 0.9676) were not statistically significant. The difference in the TN decisions was statistically significant (t = 1.926, p = 0.05). 8-bit displays will not impact negatively diagnostic accuracy, but using 11-bit displays may improve workflow efficiency.

Six observers at the site using the system C displays viewed the set of 20 images (15 with nodules, 5 without) as their eye position was recorded. Figure 4 shows a typical search pattern of one of the radiologists in the study. The circles represent fixations or locations where the eye lands with foveal (high resolution) vision. The size of the circle represents shorter (smaller) or longer (larger) dwell times. The lines connecting the circles represent saccades, or jumps, between the fixations and indicate the order in which the fixations were generated. The arrow in Fig. 4 indicates the location of the nodule. In this case, the radiologist fixated the nodule and reported it correctly (a true positive decision).

p37a_tif

FIGURE 4 — Typical scanning pattern of a radiologist searching for pulmonary nodules. The arrow indicates the location of the nodule.

Computer-vision determination of 3-D geometric parameters of LDL particles via cryogenic transmission electron microscopy

W. Lewis Collier
Lenore M. Martin
Rik van Antwerpen

Capstone Visual Product Development

Abstract — Previous research has shown that the size of the LDL macromolecules can have an effect on cardiovascular health and that low-density lipoprotein (LDL) macromolecules may be non-spherical in shape. Some of these studies, however, used methods that are not conducive to automatic determination of the 3-D parameters of the particles. In particular, the prior methods used for determination of geometric-parameter determination were either centrifugal separations or manual determination of parameters from cryogenic transmission electron micrographs. An application of computer-vision techniques to automatically determine the 3-D parameters from cryogenic transmission electron microscopy (CTEM) images will be described. Correlation of computer-generated geometric models to the orthonormal projection CTEM imagery were investigated to determine the applicability of finding the pertinent geometric parameters of the expected discoid shape of the LDL particles. The processing showed that the discoid shape can be verified using small-angle rotations that are more amenable to the limitations of CTEM imaging.

The goal of this project was to retrieve 3-D parameters of height, width, and wall thickness from a minimum number of 2-D images. By requiring only a small number of images to reconstruct the biological artifact, the likelihood that the sample is distorted (i.e., melted or altered) by exposure to the electron beam is also significantly reduced.

p37b_tif


FIGURE 6 — CTEM ray paths and depth of focus.

Digital imaging, spectroscopy, and liquid crystals in a hand-held, non-invasive device to determine hemoglobin concentration

Selim Suner
John McMurdy
Gregory Jay
Gregory Crawford

Rhode Island Hospital MedicalSimulation Center

Abstract — Anemia is a significant public-health concern both in the United States and throughout the world. This disorder of low hemoglobin concentration in the blood, which often lurks undetected for long periods, contributes significantly to mortality and morbidity and is a major cause of lost revenues from workforce shortfalls, particularly in developing countries where the incidence of anemia is higher due to malnutrition and parasitic disease. The gold standard for measuring hemoglobin is a blood test requiring phlebotomy and laboratory quantification. Current physical examination techniques and non-invasive adjuncts for detecting anemia are not sensitive and are subject to user variability. There is an urgent need for a hand-held device which can measure hemoglobin concentration accurately, inexpensively, and non-invasively. The evolution of a device from its early inception, which analyzed color decomposition in a digital image of the eyelid, to its final version that is a light, inexpensive, hand-held, non-invasive device utilizing liquid crystals to delineate the spectroscopic characteristics of the reflected light from the palpebral conjunctiva (the inside of the lower eyelid which contains many small blood vessels) and determine hemoglobin concentration in the circulating blood will be discussed. The future refinements required for bringing such a device to the market will also be discussed.

Liquid crystals can be created in morphologies such that they reflect narrow bands of light, and the peak reflectance wavelength of this narrow band can be tuned either thermally or by an electrical field. Thus, a compact and inexpensive spectrometer can be fabricated using only a single liquid-crystal panel and photodetector, allowing a fully integrated, handheld spectrometer used to measure reflectance spectra from the conjunctiva.

p38a_tif

FIGURE 6 — A photograph of data collection using a head-mount stabilizer and grating spectrometer with an imaging eyepiece to collect diffuse reflectance spectra from the tarsal plate region of the palpebral conjunctiva.

Visual quality of high-contrast projection screens. Part I: Visibility of screen-based artifacts and noise

Thomas G. Fiske
Louis D. Silverstein
Sue Hodgson
Andrew B. Watson

Rockwell Collins Display Systems

Abstract — The investigation of visual-quality issues for high-contrast projection screens such as those used for avionics projection displays and other applications requiring a high level of ambient light rejection will be described. Visibility of artifacts and noise generated by the optical and structural properties of a variety of high-contrast projection screens was characterized by both empirical, subjective image quality evaluations and a suite of objective screen-quality metrics. At least two of these metrics were found to provide very good correlation with the subjective assessment. These metrics are applicable to any display system where spatial-noise artifacts are an issue, such as with direct-view AMLCDs with anti-glare treatments.

Figure 4 shows the 2-D modulation spectrum of one screen sample. To obtain the modulation spectrum, the complex modulus of the real and imaginary parts of the Fourier transform at each spatial frequency was normalized by the amplitude of the DC term (i.e., the space-average luminance of the screen). The computed modulation is equivalent to the familiar Michelson contrast or contrast modulation:

p38c_tif

where Lmax and Lmin designate maximum and minimum luminance, respectively.

p38b_tif

FIGURE 4 — Two-dimensional modulation spectrum of high-contrast screen sample rear illuminated by a flat, white field and scaled to cycles per degree assuming a 63.5-cm viewing distance.

Process development and accurate low-cost characterization for OLED sealants by using a calcium test

Steffen Hergert
Max Linkor
Markus Korny
Norbert Fruehauf

University of Stuttgart

Abstract — A calcium measurement setup was built for testing encapsulation especially for OLED applications. This setup is able to measure both reflective and transmissive cells. For the characterization of sealants, a method to compare them with other sealing products will be described. This includes the use of spacers, a homogeneous surface energy, and the geometry of the sealant line. The effects of different geometries will be discussed. The setup was designed to achieve good accuracy at a very reasonable component cost, which will allow other facilities to replicate this setup. Therefore, the construction plan as well as the list of components can be downloaded from the authors' Web site.

The organic electroluminescent material as well as the cathode of an OLED display consists of a water- and oxygen-sensitive material. The calcium test allows the determination of the transmission rate by using the fact that pure calcium looks like a mirror, but Ca(OH)2 is transparent. Standard specifications for OLED encapsulation are on the order of 1 x 10–6 g/m2/day for water and 1 x 10–3 g/m2/day for oxygen. The existing methods need approximately 5 days or more to measure such a low transmission rate which slows down the research process. The authors believe that their setup drastically reduces the time to classify encapsulations and to speed up the research process whereas the costs of this setup are quite low because only off-the-shelf products are used.

p39_tif

FIGURE 7 — Calcium-measurement setup.